High performance Liquid chromatography(HPLC)


1.     Introduction
High-Performance Liquid Chromatography(HPLC: formerly referrd to ashigh-pressour liquid Chrometography), is a techique in analytical chemistry used to separate, identify, and quantify each component in a mixture.
The principle of chrometography, in chromatography a liquid is pumped through a bed of particles. The liquid is called the mobile phase and the particle the staionary phase . High performance liquid chrometography high performance liquid chrometography (HPLC) is basicallly a highy improved from liquid chrometography. Instead of solvent being chrometograhy allowed todrip though a column under gravity, it is forced through under high pressur of up to 400 atmospheres.That makes it much faster. All chrometograhy separation , includinHPLC operate under the sane basic principle; sepration of a samll into it’s  constituent parts because of the difference in the relative affinities of defferent molecules for the mobile phase and the stationary phase used in the sepration.
2.     Instrumentation of  HPLC





3.     HPLC Tuoubleshooting
A.   Peak Tailing
Possible Cause
Solution
1.     Blocked frit
1.      
a.     Reverse flush column ( if allowed)
b.     Replace inlet frit
c.      Replace Column

2.     Column Void
2.     Fill void
     3.  Interfering peak  
     3. 
               a.  Use longer column
               b. Change mobile phase                 and / or column/ selectivity
     4.Wrong mobile phase pH
     4.
              a. Adjust pH
              b. For basic compounds,  lower pH usually provide more symmetric peak.
4.     Sample reacting with active site
5.      
a.     Add ion pair reagent or volatile basic modifier
b.     Change column



B.   Peak Pronting


Possible Cause
Solution
1.     Low temperature
1.Increase cloumn temerature
2.     Wrong sample solvent
2.Use mobile phase for injection solvent
3.     Sample overload
3.Decrase sample concentration
4.     Bad Column
4.See A.1 and A.2

C.   Split Peaks


Possible Cause
Solution
1.     Contamination on guard or analytical column inlet
          Fig. Split Peaks



         1.
              a.Remove guard column and attempt analysis.
              b.Replace guard if necessary
              c.If analytical column is obstructed, reverse and flush
             d.If problam persists, column may be fouled with strongly reatined contaminats
             e.Use appropriate restration procedure
             f.If problam persists, inlet is probably plugged
            g.Change frit or replace column
2.     Sample solvent incompatible with mobile phase
2.Change solvent; whenever possible, inject samples in mobile phase


D.   Distortion of Larger peaks


Possible Cause
Solution
1.     Sample overload
1.Reduce sample size


E.   Distoration of Early Peaks

Possible Cause
Solution
1.     Wrong injection
1.
a. Reduce injection volume
b. Use weaker injection solvent


F.    Extra peaks

Possible Cause
Solution
1.     Other components in sample
1.Normal
2.     Late- eluting peak from previous injection
2.
a. Increase run time or gradient slope
b. Increase flow rate
3.     Vacancy or ghost peaks
3.
a. cheak purity of mobile phase
b. Use mobile phase as injection solvent
c. Reduce injection volume


G.  Retention Time Drifts

Possible Cause
Solution
1.     Poor temperature control
1.Thermotat column
2.     Mobile phase changing
2.Prevent change (evaporation, reaction.)


H.  Abrupt Retention Time Change

Possible Cause
Solution
1.     Flow rate change
 1.Reset flow rate
2.     Air bubble in pump
 2.Bleed air from pump
3.     Improper mobile phase
      3.
      a. Replace with proper mobile phase
          b.Set proper mobile phase mixture on
4.     Weak detectour lamp
       4.Replace lamp
5.     Column leaking silica or packing material
       5.Replace Column
6.     Mobile phase mixture inadequate or malfunctioning
7.     Repair or replace the mix offline if isocratic


I.      Broad Peaks

Possible Cause
Solution
1.     Mobile phase composition changed
1.Prepare new mobile phase
2.     Mobile-phase flow rate too low
           2.Adjust flow rate
3.     Leaks
           3.
         a. See Section 3
         b. Cheak for loose fittings
         c. Cheak pump for leaks, salt build- up , and unsual noises
         d. Change seals if necessary
4.     Detector setting incorrect
           4.Adjust setting
5.     Extra-column effect:
a.     Column overloaded
b.     Detectore response time oe call volume too large
c.      Tubing between column and dectector too long or ID too larg
d.     Recorder response time too high
           5.
a.     Inject smaller column (e.g. 10µl vs. 100 µl) or 1:100 and 1:100 dilution of sample
b.     Reduce response time oe use smaller call
c.      Use as short a piece of 0.007-0..10. inch ID tubing as practical
d.     Reduce respose time

6.     Buffer concentration too low
6.Increase concentration
7.     Guard column contaminated / worn out
           7.Replace guard colunm
8.     Column contaminateds / worn out: low plate number

9.     Void at column inlet







Comming Soon............................

Drx. Atul Jaiswal , Sourabh Kushwah
 and Vishal Kushwah

Documents Required for Fresh registration as pharmacist

          14. 3200/- for online registration charge.